Acquisition - trig_delay, trdl, Range, Command syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual

Page 30: Query syntax, Response format, Example, Related commands

Advertising
background image

WaveAce Remote Control

ACQUISITION - TRIG_DELAY, TRDL

The TRIG_DELAY command sets the time at which the trigger is to occur with respect to the nominal

zero delay position, which defaults to the center of the grid.

RANGE

Negative delay

: 0 to - divisions x Time/div supported by scope

Postive delay

: 0 to + division x Time/div supported by scope

If a value outside these ranges is specified, the trigger time is set to the nearest limit.

Command Syntax

TRig_DeLay <value>

<value>:=

delay time as per RANGE previously described in this topic.

NOTE: The value parameter is optional.

Query Syntax

TRig_DeLay?

Response Format

The response to the TRIG_DELAY? query indicates the trigger time with respect to the first acquired

data point.

TRig_DeLay <value>

Example

The following instruction sets the trigger delay to -20 S (post-trigger):

TRDL -20S

Related Commands

TIME_DIV, TRIG_COUPLING, TRIG_LEVEL, TRIG_MODE, TRIG_SELECT, TRIG_SLOPE