Acquisition - trig_delay, trdl, Range, Command syntax – Teledyne LeCroy WaveAce 1000_2000 Remote Control User Manual
Page 30: Query syntax, Response format, Example, Related commands
WaveAce Remote Control
ACQUISITION - TRIG_DELAY, TRDL
The TRIG_DELAY command sets the time at which the trigger is to occur with respect to the nominal
zero delay position, which defaults to the center of the grid.
RANGE
Negative delay
: 0 to - divisions x Time/div supported by scope
Postive delay
: 0 to + division x Time/div supported by scope
If a value outside these ranges is specified, the trigger time is set to the nearest limit.
Command Syntax
TRig_DeLay <value>
<value>:=
delay time as per RANGE previously described in this topic.
NOTE: The value parameter is optional.
Query Syntax
TRig_DeLay?
Response Format
The response to the TRIG_DELAY? query indicates the trigger time with respect to the first acquired
data point.
TRig_DeLay <value>
Example
The following instruction sets the trigger delay to -20 S (post-trigger):
TRDL -20S
Related Commands
TIME_DIV, TRIG_COUPLING, TRIG_LEVEL, TRIG_MODE, TRIG_SELECT, TRIG_SLOPE