STATUS - *ESR?

The *ESR? query reads and clears the contents of the Event Status Register (ESR). The response

represents the sum of the binary values of the register bits 0 to 7. The following table provides an

overview of the ESR register structure.

Query Syntax

*ESR?

Response Format

*ESR <value>

<value>:= 0 to 255

Response message:

*ESR 0

Related Commands

ALL_STATUS, *CLS, *ESE

ESR? Status Register Structure

Bit

Value

Name

Description

Note

15...8

0 - Reserved by IEEE 488.2.

7

128

PON

1 - Power OFF-to-ON

transition has occurred.

The Power On (PON) bit is always turned on (1) when the

unit is powered up.

6

64

URQ

1 - Not used.

On older Teledyne LeCroy oscilloscopes, this bit reports soft

key inputs. It does not apply to WaveAce oscilloscopes.

5

32

CME

1 - CoMmand parser Error

has been detected.

The CoMmand parser Error bit (CME) is set true (1)

whenever a command syntax error is detected. The CME bit

has an associated CoMmand parser Register (CMR) which

specifies the error code. Refer to the query CMR? for

further details.

4

16

EXE

1 - EXecution Error

detected.

The EXecution Error bit (EXE) is set true (1) when a

command cannot be executed due to some device

condition (for example, the oscilloscope in local state) or a

semantic error. The EXE bit has an associated Execution

Error Register (EXR) that specifies the error code. Refer to

query EXR? for further details.