Current Measurement (Optional)
168
Inrush pass/fail
The Inrush pass/fail is determined by the area under the curve and above the
Y=100 mA horizontal line. If the area is less than 50 uC, the test is reported as
a pass.
Figure 129 Inrush Current Display With Cursors
Enabling the cursors allows you to make precise measurements in the display.
The time and current values at the intersection of the X(Time) and Y(Current)
cursors are displayed on the toolbar as well as the area under the curve
representing the energy in micro coulombs.
Positioning the Cursors
To position the cursors within the display, click the pointer at the point to place
a cursor.