RADWAG WPW 60/H5/K Multifunctional Scales User Manual
Page 122
Advertising
Press
to initialise COUNTING PCS. with automatic setting
of standard mass.
Caution:
Single piece mass cannot be greater than the full range of the main
platform and less than 0.1 of reading division (d) of the scale.
24.8.14. Setting the reference standard mass by declaring the
quantity of the standard
Procedure:
• Enter the COUNTING PIECES mode according
to chapter 24.8.12 of this manual:
- 122 -