By entering a standard mass from the database – RADWAG WPW 60/H5/K Multifunctional Scales User Manual
Page 124
There are four conditions of working of Automatic Accuracy
Correction:
1. quantity of pieces should be increased
2. addend quantity of pieces should not be greater than double number
of pieces on the pan.
3. The new sample can be different from the old sample by
± 0.3 of pcs
(absolute value),
Example:
First weighing:
50 pcs, 1 pcs=0.5g, total=25g
Second weighing:
50+30 pcs, total=39.7g
It means that having a unit mass 0.5g we have 79.4 pcs. The expected
value was 80 pcs. The difference is 0.6 that is greater than 0.3 and
Auto-correction cannot work.
4. equilibrium should be reach.
If a user recognises the reference quantity to be adequate, he can save the
unit mass (see 25.4). At the same time the function of Automatic Correction
of Accuracy will be switched off what is signalled by continuous display of
the
pictogram. Users can also stop the operation of the function by
pressing
.
Caution:
Automatic Accuracy Correction works only for main scale.
Return to weighing:
See 11.3.
24.8.15. By entering a standard mass from the database
After selecting a product from the assortment database, the unit mass
is ascribed.
Procedure:
While in counting pieces press
- assortment database.
Select a required product (
,
) and confirm with
.