29. ERROR MESSAGES


ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING
(where: XXX – parameter name) – confirmed with ENTER
of a wrong value in the user’s menu,

Err2

- Value beyond the zero range,

Err3

- Value beyond the tare range,

Err4

- Calibration mass or start mass adjustment error

(

±1% for weight, ±10% for start mass),

Err5

- Mass of a single piece lower than the 0.1 of scale division,

Err6

- Mass of all pcs (declared as standard) lower than one

division,

Err8

- Tarring / zeroing operation time exceeded,

NULL

- Zero value from the AD converter,

FULL2

- Measurement

range overflow,

HI

- Display range overflow,

LH

- Start mass error, the mass on the weighing platform

is beyond the acceptable range

±20 of start mass



30. TROUBLE SHOOTING

Problem Cause

Solution

Scale switches off
automatically

„t1” parameter set to „YES”
(Power save)

In <P9 OTHERS> change
<P9.3 AUTO SWITCH.> to
„NO”

„LH” appears on the
display

Weight pan not empty
when switching on

Unload the pan. Indication
returns to zero