Lock the scale keyboard, Unlock the scale keyboard, Initiating of dosing/filling – RADWAG WPW 60/H5/K Multifunctional Scales User Manual
Page 163: Stop of dosing/filling, Set lower threshold
Advertising
- 163 -
28.4.14. Lock the scale keyboard
Syntax: K1 CR LF
Possible answers:
K1_I CR LF
- command comprehended but cannot be executed
K1_OK CR LF
- command executed
Caution:
This command is not remembered after restart
28.4.15. Unlock the scale keyboard
Syntax: K0 CR LF
Possible answers: K0_OK CR LF – command in progress
28.4.16. Initiating of dosing/filling
Syntax: S1 CR LF
Possible answers:
S1_I CR LF
- command comprehended but cannot be executed
S1_OK CR LF
- command in progress
28.4.17. Stop of dosing/filling
Syntax: S0 CR LF
Possible answers:
S0_I CR LF
- command comprehended but cannot be executed
S0_OK CR LF
- command in progress
28.4.18. Set lower threshold
Syntax: DH_XXXXX CR LF, where: XXXXX – mass format