Send all implemented commands, Manual printouts / automatic printouts – RADWAG WPW 60/H5/K Multifunctional Scales User Manual
Page 165
Mass - 9 characters justified to the right
Unit
- 3 characters justified to the left
28.4.22. Send all implemented commands
Syntax: PC CR LF
Possible answers:
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,PC,K1,K0,DH,UH,
ODH,OUH,S1,S0,OT,UT" – command executed, the indicator
have sent all the implemented commands.
28.5. Manual printouts / automatic printouts
Users can general manual or automatic printouts from the scale.
• Manual printouts can be performed after loading the pan and
stabilizing indication by pressing
.
• Automatic printouts can be performed only after loading the
pan and stabilizing indication.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1
2
3 4
-12 13 14 15 16 17 18
stability space
sign
mass space
unit
CR LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
sign
[space] for positive values or
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
- 165 -