Send all implemented commands, Manual printouts / automatic printouts – RADWAG WPW 60/H5/K Multifunctional Scales User Manual

Page 165

Advertising
background image

Mass - 9 characters justified to the right
Unit

- 3 characters justified to the left



28.4.22. Send all implemented commands

Syntax: PC CR LF

Possible answers:

PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,PC,K1,K0,DH,UH,
ODH,OUH,S1,S0,OT,UT"
– command executed, the indicator
have sent all the implemented commands.



28.5. Manual printouts / automatic printouts

Users can general manual or automatic printouts from the scale.

• Manual printouts can be performed after loading the pan and

stabilizing indication by pressing

.

• Automatic printouts can be performed only after loading the

pan and stabilizing indication.

Notice:
If a scale is verified printouts of immediate values are blocked.

Format frame:

1

2

3 4

-12 13 14 15 16 17 18

stability space

sign

mass space

unit

CR LF

Stability character

[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range

sign

[space] for positive values or
[-] for negative values

mass

9 characters justified to the right

unit

3 characters justified to the left

command

3 characters justified to the left

- 165 -