Send the stable result in current unit – RADWAG WPW 60/H5/K Multifunctional Scales User Manual

Page 160

Advertising
background image

- 160 -

n

- weighing platform number

mass - 9 characters justified to the right
unit - 3 characters justified to the left


Example:

Let us assume that both platforms are connected to indicator PUE C41H.

S I A CR LF – computer command
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF
- command done, mass values
from both platforms are returned in basic units



28.4.8. Send the stable result in current unit

Syntax: SU CR LF

Possible answers:


SU_A CR LF
SU_E CR LF

- command accepted and in progress
- timeout while waiting for stable results

SU_I CR LF

- command comprehended but cannot be executed

SU_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in current unit is returned


Frame format:

1 2

3

4

5

6 7-15 16 17 18 19 20 21

S U space stability space sign mass

space

unit

CR

LF


Example:

S U CR LF – computer command
S U _ A CR LF - command accepted and in progress

S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass
value in current unit is returned.