7. START UP

First unpack and assemble the scale (see point 5 of this user manual).
Then:

• Level the scale using its adjustable feet located in the bottom base of

the medical scale. Monitor the level status on the level located in the
base of the medical scale. Leveling is correct if the air bubble
of the level indicator is located in the central location:

• Switch on the scale by pressing

key – press and hold the key for

0,5 second,

• On turning on wait until the scale completes the auto test procedure,

• Then the zero mass indication appears on the display, with the

following pictograms:

- precise zero indication

- stable measurement result

kg

- measuring unit

• If measurement result is other than zero, press zeroing key on

indicator’s overlay.