20.5. Manual printout

Medical scale enables generating manual printouts. The manual printout is
generated after placing a weighed load on scale’s weighing platform,

stabilization of measurement result and pressing

key.

Caution:
If the scale is a verified one, then printout of unstable / instantaneous data
in disabled.

Format of mass printout:

1

2

3

4 -12

13

14

15

16

17

18

stability

marker

space

sign

mass

space

unit

CR

LF

Stability marker

[space] if stable
[?] if unstable
[^] if error of exceeding range to “+” occurs
[v] if error of exceeding range to “-“ occurs

sign

[space] for positive values or
[-] for negative values

mass

9 characters with decimal point and right

justification

unit

3 characters with left justification

command

3 characters with left justification


Example 1:

_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.


Example 2:

? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.


Example 3:

^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - a printout generated from a scale on
pressing ENTER/PRINT key.