Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
marker
space
sign
mass
space
unit
CR
LF
Example:
S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response of mass value in basic weighing unit
20.4.6. Send stable measurement result in current weighing unit
Format: SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement result
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
space
stability
marker
space
sign
mass
space
unit
CR
LF
Example:
S U CR LF – command from a computer
S U _ A CR LF – command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.