Continuous transmission, Printout configuration, Error messages – RADWAG WPT/8B 300C Ramps Bed Scales User Manual
Page 54
- 54 -
20.6. Continuous transmission
Scale indicator enables printing mass value in continuous transmission in
basic and in current measuring unit. Continuous transmission mode in
medical scales can be enabled by sending a command via RS 232C
interface (see point 20.4 of this user manual).
20.7. Printout configuration
General information
If data contained on a standard printout is excessive or insufficient, and
there is a need to modify it according to user needs, please use SCALE
EDITOR computer software. The freeware is available for downloading
from RADWAG website: http://www.radwag.com
21. ERROR MESSAGES
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Adjustment mass or start mass beyond the
acceptable range (
±1% for weight, ±10 for start mass)
Err5
- Mass of single part (piece) below scale’s reading unit.
Err7
- Too short scale’s switch off time (recommended time
more than 3 sec.)
Err8
- Tarring / Zeroing operation time exceeded
null
- Zero value from the AD converter
FULL2
- Measurement range (Max. capacity) exceeded
LH
- Start mass error, indication beyond acceptable range
(from -5% to +15% of start mass)
Caution:
1. Errors: Err2, Err3, Err4, Err8, null, LH, that appear on the display are
also accompanied by a short beep sound (about 1 sec.).
2. Error FULL2 that appears on the display is also accompanied by
a continuous sound until the cause of error (excessive load on the
weighing platform) disappears.