25. EMC INFORMATION

Table 201

Manufacturer declaration and information – electromagnetic emission

The DEVICE is intended to operate in the electromagnetic environment described below.

Recipient or user of the DEVICE should make sure that it operates in such an environment.

Emissivity tests

Compliance Electromagnetic environment – information

Emission of RF

according to CISPR 11

Group 1

The DEVICE produces energy in the range of RF only

as a subsequence of internal functionality. Owing to

that the emission in the range of RF is minimal and it

is unlikely for it to interfere with anything in the near

proximity.

Emission in the range

of RF according to

CISPR 11

Class B

The DEVICE is relevant to operate in all premises,

including living surface and premises directly

connected to the low-tension system (mains) that

supplies dwelling houses.

RF – frequency from the interval o electromagnetic spectrum which is between the lower

range of radio frequencies of long waves and the infrared range; frequencies useful for radio

transmission. Limit frequencies are 9 kHz and 3 000 GHz

Table 202

Manufacturer declaration and information - electromagnetic immunity

The DEVICE is intended to operate in the electromagnetic environment described below.

Recipient or user of the DEVICE should make sure that it operates in such an environment.

Immunity tests

Testing level IEC

60601

Met requirements Electromagnetic

environment – information

Electrostatic

discharge (ESD)

according to IEC

61000-4-2

+/- 6kV contact

discharge

+/- 8kV through-air

discharge

+/- 6kV contact

discharge

+/- 8kV through-air

discharge

The floor should be wooden,

concrete or made of ceramic

plates. If floors are covered

with a synthetic material

relative humidity should be

at least 30%.

Transient states

and impulses

according to IEC

61000-4-5

+/- 2 kV

for supply lines

+/- 1 kV

for I/O lines

+/- 2 kV

for supply lines

+/- 1 kV

for I/O lines

The quality of mains should

be on the on the office- or

hospital-level environment.

Surge

disturbances

according to IEC

61000-4-5

+/- 1 kV

differential mode

+/- 2 kV

associative mode

+/- 1 kV

differential mode

+/- 2 kV

associative mode

The quality of mains should

be on the on the office- or

hospital-level environment.