Determining bmi factor – RADWAG WPT/8B 300C Ramps Bed Scales User Manual

Page 39

Advertising
background image

- 39 -


Caution:
In case the mass indication is below set value of S_Lo parameter, then the
measurement result will not be locked (HoLd function will not be activated).

Mode operation:

• Load scale’s weighing platform. On stabilization of measurement result

the display locks mass indication, as required by enabled HoLd mode.

• Take the weighed load off scale’s weighing platform,

• Before carrying out the following measurement press

key.



16.3. Determining BMI factor

The mode of determining body mass index BMI is dedicated for adults over
18 years old, therefore the software comprises the following limitations:

• The mode is not available in baby scales (models WPT 6/15D and

WPT 10/20D).

• Body mass limitation for mode activation - min 10kg,
• Height limitation for mode activation - from 1m to 2,5m.


Mode operation:

• Load scale’s weighing platform,

• Press function key

to select working mode bnni,

• If measurement result is stable, the display indicates (for

approximately 2 seconds) the mass value of weighed patient followed
by blinking value of height expressed in meters,

• If measurement result is unstable, the display indicates dashes

< - - - - - - - - - - > until stable measurement result is obtained. Then
the display indicates (for approximately 2 seconds) the mass value of
weighed patient followed by blinking value of height expressed in
meters.

Advertising