Frame format, Format: s cr lf, Accessible responses – RADWAG WPT/8B 300C Ramps Bed Scales User Manual
Page 49
- 49 -
Frame format:
1
2
3
4
5-6
7-15
16
17
18
19
20
21
T
O
space
stability
marker
space
tare
space
unit
CR
LF
Tare - 9 characters with right justification
Unit
- 3 characters with left justification
20.4.4. Send stable measurement result in basic measuring unit
Format: S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Frame format:
1
2-3
4
5
6
7-15
16
17
18
19
20
21
S
space
stability
marker
space
sign
mass
space
unit
CR
LF
Example:
S CR LF – command sent from a computer
S _ A CR LF - command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response
is mass value in basic measuring unit.
20.4.5. Immediately send measurement result in basic measuring unit
Format: SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response is immediate with mass value in basic weighing unit