Send all implemented commands, Format: c0 cr lf, Accessible responses – RADWAG WPT/8B 300C Ramps Bed Scales User Manual

Page 52: Format: cu1 cr lf, Frame format, Format: cu0 cr lf, Send all implemented commands format: pc cr lf

Advertising
background image

- 52 -

20.4.9. Switch off continuous transmission in basic measuring unit

Format: C0 CR LF

Accessible responses:

C0_I CR LF

- command understood but not accessible at this moment

C0_A CR LF

- command understood and in progress


20.4.10. Switch on continuous transmission in current measuring unit

Format: CU1 CR LF

Accessible responses:

CU1_I CR LF

- command understood but not accessible at this moment

CU1_A CR LF
MASS FRAME

- command understood and in progress
- response is mass value in current measuring unit


Frame format:

1

2

3

4

5

6

7-15

16

17

18

19

20

21

S

U

I

stability

marker

space

sign

mass

space

unit

CR

LF


20.4.11. Switch off continuous transmission in current measuring unit

Format: CU0 CR LF

Accessible responses:


CU0_I CR LF

- command understood but not accessible at this moment

CU0_A CR LF

- command understood and in progress


20.4.12. Send all implemented commands

Format: PC CR LF

Accessible responses:

PC_- >_Z,T, TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command carried
out, terminal has sent all implemented commands.