Trouble shooting, Additional equipment – RADWAG WPT/8B 300C Ramps Bed Scales User Manual

Page 57

Advertising
background image

- 57 -

23. TROUBLE SHOOTING

Problem

Cause

Solution

Scale does not
turn on

Discharged accumulator /
accumulators,

Connect power adapter to
charge the accumulator /
accumulators.

No accumulators (not
installed or improperly
installed)

Check the correctness of
accumulator(s) installation
(polarization).

Scale turns off
automatically

Parameter “t1” set to “YES”
(power saving mode)

Go to submenu “othr” and
change parameter
“5.4 t1” to “no”.

After turning on
message “LH” is
displayed

Load on scale’s weighing
platform during powering up.

Unload the weighing platform.
In short period of time the
scale should indicate zero.



24. ADDITIONAL EQUIPMENT


Accessories:

• KAFKA printer cable - P0136,
• Computer cable - P0108,

• EPSON printer cable - P0151,

• Power cord for car lighter 12V DC - K0047,
• Thermal printer - KAFKA,

• Dot matrix printer - EPSON,
• Additional display in plastic housing - WD- 4/1

(available only as the factory installed option with a new scale),

• RS232 / RS485 converter for PUE C/31 – KR-01,

• RS232 / Ethernet converter for PUE C/31 - KR-04.


Computer software:

• „Scale Editor” computer software,

• "RAD-KEY" computer software,
• "PW-WIN" computer software.

Advertising