Scale adjustment, Adjustment – RADWAG WPT/8B 300C Ramps Bed Scales User Manual
Page 41
- 41 -
17. SCALE ADJUSTMENT
Available in non-verified scales only
Ensuring high weighing accuracy requires periodical controlling scale’s
indications and correcting the adjustment factor stored in scale’s memory –
the process is known as scale adjustment.
Adjustment should be carried out on scale start of operation or if dynamic
change of temperature occurs. Before carrying out adjustment procedure
remove load from the weighing platform.
17.1. Adjustment
Procedure:
• Enter submenu <P6.CAL> in accordance with point 11.2 of the user
manual, and:
• The software displays the following messages / commands:
• During this time the scale adjusts its start mass, and after completing
the procedure, the mass of adjustment weight is
displayed (e.g. 3.000kg).