- 36 -

CHr6 YES

- function enabled (pictogram appears on the

display)

CHr6 no

- function disabled (pictogram does not appear on

the display)


Return to weighing:

See point - 11.2.2. – return to weighing.



15.4.4. Accumulator charging process

The indicator of a medical scale comes standard with a set of new NiMH
accumulators, size AA (R6), and a power adapter. On scale’s
commissioning it is obligatory to form the accumulators, which extends their
life-span. The process requires full charging and complete discharging the
accumulators. New accumulators have the optimum capacity after
completing three charging / discharging cycles.

Forming process:

1. Plug the power adapter to scale’s indicator.
2. Charge the accumulators for approximately 12 hours – charging time

specified for accumulators with capacity 2200mAh.

3. After 12 hours unplug the power adapter from the indicator.
4. Operate the scale with supply from accumulators until its automatic

switches off.

Advertising