9 error cindition, 9 error condition – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 104

Advertising
background image

105

Figure 4-34

4.4.9 Error Condition

The error annunciator on the front panel display tells about the error

condition of the multimeter. If there are one or more syntax or hardware

errors found, the error annunciator will be lit to inform user. The

multimeter stores errors using the first-in-first-out (FIFO) order and it

records up to 20 errors in the error queue. The first error detected will be

the first error retrieved. Refer to Chapter 6 for the list of error

messages.

1. The error annunciator will be turned off when you have read the

errors in the memory.

2. If there is no error found when you check the error queue, the

multimeter shows “NO ERRORS”.

3. If there are more than 20 errors found, the last error in the error

queue will be replaced with “-350”, means too many errors.

4. The error queue is cleared when the multimeter is turned off, or when

a *CLS command is received.

How to check the error queue

Press MENU then use ◁ and ▷ to locate “SYSTEM” submenu. Press

ENTER to select it. Again use ◁ and ▷ to locate “ERROR” option. Press

ENTER to select it. The message about error queue will show on the

display. Refer to Chapter 6 for the descriptions of all the error

messages. The locations of the buttons are shown with red rectangle

frames in Figure 4-35.

Advertising