3 trigger setting – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 73

Advertising
background image

74

number of samples per trigger, the number of triggers per event, reading

hold, and the trigger delay for their measurements.

A. Number of samples on each trigger

By default, the multimeter takes only one reading on each trigger, but

you can instruct the multimeter to take specific number (up to 50000) of

readings each time it receives a trigger. The user’s input setting is stored

in a volatile memory which will be cleared after the meter has been

turned off and the default value will be restored. You can set the number

of samples on each trigger through front panel or the remote interface.

Front Panel Operation

Press MENU button and locate “TRIG” submenu by pressing ◁ and ▷

buttons. Press ENTER to choose “TRIG” option. Again locate “N SAMPLE”

submenu using ◁ and ▷ buttons and then press ENTER to select it.

Then set the desired number by pressing △and ▽ buttons to increase

or decrease the number, and using ◁ and ▷ buttons to move through

the digits. The locations of these buttons are shown with red rectangle

frames in Figure 4-11.

Procedures: MENU TRIG N SAMPLE <Number>

Figure 4-11

Remote Interface Operation

Use the following command to set the number of samples from your PC

terminal.

+