Front Panel Operations

This chapter contains information about how to change the parameters

and settings for your measurements and all the details about each

feature and function.

4.1 Measurement Configuration

The following information will guide you through ways to configure

measurement functions. It provides you the flexibility to change any

parameter in any measurement function when needed, including ADC

setting, filter, resolution setting (digits), DC input resistance, threshold

resistance (continuity), range (manual & auto), rate (integration time),

sensor type for temperature measurements, remote interface selection

and input terminal switch.

4.1.1 Set ADC (Auto Zero & Auto Gain)

Zero and Gain

The purpose of Auto Zero and Auto Gain functions are used for

minimizing the offset influence on your measurements. When Auto Zero

or Auto Gain is enabled, 1201 takes the input signal reading as a base

value and then internally disconnects the input signal, and takes a offset

reading (a null offset). It then subtracts the offset from the base to get

an accurate measurement.

Displayed reading = Base value (input signal) – Offset value

When the Auto Zero or Auto Gain is enabled the meter takes an offset

reading for each subsequent measurement. However, when Auto Zero or

Auto Gain is disabled, the meter only takes one offset reading each time

when you change the function settings. User can set “PERIOD” to change

the time interval for taking an offset reading.