4 dc input resistance – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 59

Advertising
background image

60

Remote Interface Operation

Use the following commands on your PC terminal to set the resolution for

your measurement.

CONFigure:<function> <range>,<resolution>

MEASure:<function>? <range>,<resolution>

SENSe:<function>:RESolution <resolution>

4.1.4 DC Input Resistance

Definition

To reduce the effect of loading errors due to the input resistance, 1201

allows user to select a much larger input resistance (> 10G Ω) for low

input DC voltage (100mV, 1V and 10V) measurements. This feature is

only available for DC voltage measurements and it is not applicable to

other measurement functions.

Default

The default input resistance for all measurements is 10M Ω. Please refer

to Table 2-1 on page 24. The DC input resistance can only be changed

for measurements with 100mV, 1V or 10V range. For 100V DCV, 1000 V

DCV and other measurement functions, the input resistance is fixed at

10M Ω and can not be changed. Also note that the user selected value

is stored in a volatile memory. The default setting will be restored after

the meter has been turned off.

How to set the DC input resistance

You can set the DC input resistance either through the front panel

operation or through the remote interface operation.

Front Panel Operation

Press CONFIG + DCV , and then use ◁ and ▷ buttons to locate “INPUT

R” option. Press ENTER to select it. Then select desired value for the

input resistance by pressing ◁ or ▷ button. Press ENTER to choose it.

Two values are available: 10M Ω or > 10G Ω. The locations of these

buttons are shown with red rectangle frames in Figure 4-4.