BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 76

Advertising
background image

77

Table 4-4

How to specify a delay time

User can set the delay time from either the front panel operations or the

remote interface operations.

Front Panel Operation

To set the delay time, press MENU button and use ◁ and ▷ buttons to

Measurement Function

Setting

Trigger Delay

Time

PLC >= 1

1.5 ms

DCV/DCI

PLC < 1

1.0 ms

100Ω ~ 100kΩ

1.5 ms

1 MΩ

15 ms

Ω2 and Ω4

(PLC >= 1)

10 MΩ ~ 100 MΩ

100 ms

100Ω ~ 100kΩ

1.0 ms

1 MΩ

10 ms

Ω2 and Ω4

(PLC < 1)

10 MΩ ~ 100 MΩ

100 ms

3 Hz

7.0 s

20 Hz

1.0 s

ACV/ACI

(Remote Interface/

External Trigger/

Single Trigger)

200 Hz

600 ms

3 Hz

1.5 s

20 Hz

200 ms

ACV/ACI

(Front Panel w/ Auto

Trigger On)

200 Hz

100 ms

Remote Interface /

External

1.0 s

Frequency/Period

Front Panel w/ Auto

Trigger ON

0 s

Advertising