BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 155

Advertising
background image

156

Set a trigger delay time in seconds. The delay is the time between the trigger

signal and each sample that follows. Specify a delay time from 0 to 3600

seconds.

TRIGger:DELay?

Query the trigger delay time.

TRIGger:DELay:AUTO {OFF|ON}

Disable or enable a automatic trigger delay. The delay is determined by function,

range, integration time, and ac filter setting. Specifying a delay time

automatically turns off the automatic trigger delay.

TRIGger:DELay:AUTO?

Query the automatic trigger delay mode. Returns “0” (OFF) or “1” (ON).

SAMPle:COUNt {<value>|MINimum|MAXimum}

Set the number of readings (samples) the multimeter takes per trigger. Select a

number from 1 to 50,000 readings per trigger.

SAMPle:COUNt ? [MINimum|MAXimum]

Query the sample count.

TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite}

Specify the number of triggers the multimeter will accept before returning to the

“idle” state. Select from 1 to 50,000 triggers. The INFinite parameter instructs

the multimeter to take triggers continuously. Trigger count is ignored while in

local operation.

TRIGger:COUNt? [MINimum|MAXimum|INFinite]

Query the trigger count. Returns “9.90000000E+37” when you select the

infinite trigger count.

C.8 The System-Related Commands

FETCh?

Transfer readings stored in memory by the INITiate command to output buffer