BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 107

Advertising
background image

108

Warning! Erroneous self-test failures may occur if the setting of

power line voltage is incorrect. Errors may also occur because of signals

present on the input terminal (front and rear) during self-test procedure.

Long test leads can act as an antenna causing pick-up of ac signals.

How to execute Self-test

Users may execute a complete self-test by front panel operation. This

test procedure provides more tests for the hardware of 1201 than the

power-on tests. Please follow the Figure 4-38 direction and procedure

below to perform the complete self-test.

Figure 4-38

Procedure: MENU

SYSTEM

SELF TEST

After self-test procedure, the result, PASS or FAIL, will be shown. If the

result is FAIL, the “ERR” annunciator on the display panel will be lit, and

error codes will be stored. You can check the error codes by the following

procedure.

Procedure: MENU

SYSTEM

ERROR

The descriptions of self-test procedures are listed below including test

number, purpose, test setup, and failure criteria.

601 Front panel does not respond The main CPU U1601 attempts to

establish serial communications with the front panel processor U3.

Advertising