C.9 status reporting commands – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 158

Advertising
background image

159

SYSTem:ERRor?

Query the multimeter’s error queue. Up to 20 errors can be stored in the

queue. Errors are retrieved in first-in-first out (FIFO) order. Each error

string may contain up to 80 characters.

STATus:QUEStionable:ENABle <enable value>

Enable bits in the Questionable Data enable register. The selected bits are then

reported to the Status Byte.

STATus:QUEStionable:ENABle?

Query the Questionable Data enable register. The multimeter returns a

binary-weighted decimal representing the bits set in the enable register.

STATus:QUEStionable:EVENt?

Query the Questionable Data event register. The multimeter returns a

decimal value which corresponds to the binary-weighted sum of all bits

set in the register.

STATus:PRESet

Clear all bits in the Questionable Data enable register.

*CLS

Clear the Status Byte summary register and all event registers.

*ESE <enable value>

Enable bits in the Standard Event enable register. The selected bits are then

reported to the Status Byte.

*ESE?

Query the Standard Event enable register. The multimeter returns a

decimal value which corresponds to the binary-weighted sum of all bits

set in the register.

*ESR?

Query the Standard event register. The multimeter returns a decimal

value which corresponds to the binary-weighted sum of all bits set in the

register.