BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 130

Advertising
background image

131

531 Insufficient memory

There is not enough memory to store the requested number of readings

in internal memory using the INITiate command. The product of the

sample count (SAMPle:COUNt) and the trigger count (TRIGger:COUNt)

must not exceed 512 readings.

532 Cannot achieve requested resolution

The multimeter cannot achieve the requested measurement resolution.

You may have specified an invalid resolution in the CONFigure or

MEASure command.

540 Cannot use overload as math reference

The multimeter cannot store an overload reading (9.90000000E+37) as

the math reference for null or dB measurements. The math state is

turned off as a result of this condition.

550 Command not allowed in local

The multimeter received a READ? Command while in the local mode.