Speed of scanner card measurement – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 99

Advertising
background image

100

DC Signals: 110V DC, 1A switched, 30VA maximum (resistive load).
AC Signals: 125V AC rms or 175V AC peak, 100kHz maximum, 1A switched,
62.5VA maximum (resistive load).
Contact Life: >100000 operations at maximum signal level; >100000000
operations cold switching.
Contact Resistance: <1ohm at end of contact life.
Actuation Time: 5ms maximum on/off.
Contact Potential: <±500nV typical per contact, 1μV max.
<±500nV typical per contact pair, 1μV max.
Connector Type: Screw terminal, #22 AWG wire size.
Isolation Between Any Two Terminals: >10 Gohm, <75pF.
Isolation Between Any Terminal and Earth: >10 Gohm, <150pF.
Common Mode Voltage: 350V peak between any terminal and earth.
Maximum Voltage Between Any Two Terminals: 200V peak.
Maximum Voltage Between Any Terminal and 1201 Input LO: 200V peak.

ENVIRONMENTAL: Meets all 1201 environmental specifications.

1201-opt01 Scanner Card Configuration & Speed List:

Speed of Scanner Card Measurement

AutoZero OFF,AutoGain OFF,AutoRange OFF, Scan Timer=0,60Hz

NPLC

Take Time with 2000 Readings(sec)

rate(ch/s)

(Fast 4.5) 0.02

68

29.4

single

function(VDC)

(Slow 4.5 & Fast 5.5) 0.1

74

27.0