1 thermocouple measurements – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 43

Advertising
background image

44

Sensor Type

Temperature

Range(°C)

Temperature

Range(°F)

E

-270~1000

-518~1832

J

-210~1200

-140~2192

S

-50~1768

-122~3236

T

-270~400

-518~752

RTD (PT 100)

-200~850

-392~1562

Table 3-1

3.7.1 Thermocouple Measurements

Connect the thermocouple adapter to the banana jacks on the front

panel. The difference between each type is subject to the patch thermal

leads.

※ Note: Only connection via the front panel can be used for temperature

measurements.

How to measure thermocouple

Use terminals switch to select terminals on the front panel.

To measure thermocouple must be via thermocouple adapter and the

low thermal patch leads as shown in Figure 3-12. (When the

mentioned modules are well connected as the following figure, please

go to the next step.)

Configure the thermocouple type and unit using CONFIG + SHIFT +

TEMP, such as K TYPE and ºC. When ready, press ENTER button.

Advertising