CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}

Preset and configure a period measurement with the specified range and

resolution. This command does not initiate the measurement. For period

measurements, the meter uses only one “range” for all inputs between 0.33

seconds and 3.3 µsec. With no input signal applied, period measurements return “0”.

CONFigure:CONTinuity

Preset and configure for a continuity measurement. This command does not

initiate the measurement. The range and resolution are fixed at 1kΩ and 5½

digits respectively.

CONFigure:DIODe

Preset and configure for a diode measurement. This command does not initiate

the measurement. The range and resolution are fixed at 1Vdc with 1mA current

source and 5½ digits respectively.

CONFigure:TEMPerature

Preset and configure for a RTD measurement. This command does not initiate

the measurement. The resolution is fixed at 6½ digits.

CONFigure:TCOuple

Preset and configure for a thermocouple measurement. This command does not

initiate the measurement. The resolution is fixed at 6½ digits.

CONFigure?

Query a recent condition of function, and return a quoted string.

C.5 The Measurement Configuration Command

[SENSe:]FUNCtion “<function>”

Select a measurement function and enclose it in quotes in the command string

(FUNC “VOLT:DC”). Use one of the following strings.

VOLTage:DC

VOLTage:AC

VOLTage:DC:RATio

CURRent:DC

CURRent:AC

RESistance (for 2-wire ohms)

FRESistance (for 4-wire ohms)

FREQuency

PERiod

CONTinuity

DIODe TCOuple TEMPerature