C.4 the configure command – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 142

Advertising
background image

143

resolution. The reading is sent to the output buffer. For frequency

measurements, the meter uses only one “range” for all inputs between 3Hz and

300kHz. With no input signal applied, frequency measurements return “0”.

MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}

Preset and make a period measurement with the specified range and resolution.

The reading is sent to the output buffer. For period measurements, the meter

uses only one “range” for all inputs between 0.33 seconds and 3.3 μsec. With

no input signal applied, period measurements return “0”.

MEASure:CONTinuity?

Preset and make a continuity measurement. The reading is sent to the output

buffer. The range and resolution are fixed at 1kΩ and 5 ½ digits respectively.

MEASure:DIODe?

Preset and make a diode measurement. The reading is sent to the output buffer.

The range and resolution are fixed at 1Vdc with 1mA current source and 5 ½

digits respectively.

MEASure:TEMPerature?

Preset and make a temperature measurement with RTD. The reading is sent to

the output buffer.

MEASure:TCOuple?

Preset and make a temperature measurement with thermocouple. The reading

is sent to the output buffer.

C.4 The CONFigure Command

CONFigure:VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}

Preset and configure the multimeter for DC voltage measurements with the

specified range and resolution. This command does not initiate the

measurement.

CONFigure:VOLTage:DC:RATio {<range>|MIN|MAX|DEF },{<resolution>|MIN|MAX|DEF}

Preset and configure the multimeter for DC:DC ratio measurements with the

specified range and resolution. This command does not initiate the