Function

Range

Resolution

Shunt

Resistance

1 Year

(23°C ± 5°C)

10.000000mA

10 nA

5.1Ω

0.050 + 0.020

100.00000mA

100 nA

5.1Ω

0.050 + 0.005

1.000000A

1 uA

0.1Ω

0.100 + 0.010

DCI

(DC

Current)

3.00000A

10 uA

0.1Ω

0.120 + 0.020

Function

Range

Resolution

Test Current

1 Year

(23°C ± 5°C)

100.0000 Ω

100 uΩ

1 mA

0.010 + 0.004

1.000000 KΩ

1 mΩ

1 mA

0.010 + 0.001

10.00000 KΩ

10 mΩ

100 uA

0.010 + 0.001

100.0000 KΩ

100 mΩ

10 uA

0.010 + 0.001

1.000000 MΩ

1 Ω

5 uA

0.010 + 0.001

10.00000 MΩ

10 Ω

500 nA

0.040 + 0.001

Resistance

4

100.0000 MΩ

100 Ω

500 nA||10

0.800 + 0.010

Diode Test

1.0000V

10 uV

1mA

0.010 + 0.020

Continuity

1000.00KΩ

10 mΩ

1mA

0.010 + 0.030

4 a. Specifications are for 4-wire ohms. For 2-wire ohms, use Math

Null function

b. Max. Lead Resistance: 10% of range per lead for 100Ω and 1KΩ

ranges; 1kΩ per lead for all other ranges.

c. Input protection: 1000 V, all ranges.