C.8 the system-related commands – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 156

Advertising
background image

157

where you are able to read them into your bus controller.

READ?

Change the state of the triggering system from the “idle” state to

“wait-for-trigger” state. The meter will start to make measurements when a

required triggering condition is met after the READ? command is received.

Measurement readings are sent to the output buffer immediately.

DISPlay {OFF|ON}

Turn off or on the display.

DISPlay?

Query the display setting. Returns “0” (OFF) or “1” (ON).

DISPlay:TEXT <quoted string>

Show a message on the front panel display. The allowed message can be up to

16 characters in the lower row display; any additional characters will be

truncated.

DISPlay:TEXT?

Query the message sent to the front panel display.

DISPlay:TEXT:CLEar

Clear the message shown on the front panel display.

SYSTem:BEEPer

Issue a single beeping tone.

SYSTem:BEEPer:STATe {OFF|ON}

Disable or enable the beeper.

SYSTem:BEEPer:STATe?

Query the beeper mode. Returns “0” (OFF) or “1” (ON).

SYSTem:ERRor?

Query the multimeter’s error queue. Up to 20 errors can be stored in the queue.

Errors are retrieved in first-in-first-out (FIFO) order. Each error string may

contain up to 80 characters.

Advertising