BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 53

Advertising
background image

54

Fast 5 ½ digits

Off

Off

0.1

Slow 5 ½ digits

On

On

1

Fast 6 ½ digits

On

On

1

Slow 6 ½ digits

On

On

10

Table 4-1

※ Note: To change resolution, refer to Section 4.1.3: Resolution Setting

in this chapter.

Figure 4-1

Remote Interface Operation

To set Auto Zero and Auto Gain through the remote interface, use the

following commands:

SENSe:ZERO:AUTO {OFF|ONCE|ON}

SENSe:GAIN:AUTO {OFF|ONCE|ON}

The OFF and ONCE parameters have a similar effect. Auto Zero OFF or

Auto Gain OFF doesn’t issue a new offset measurement. However, Auto

Zero ONCE or Auto Gain ONCE issues an immediate offset measurement.

Advertising