BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 75

Advertising
background image

76

system is called the “settling time.” The settling time depends on the

measurement range, cable properties and signal source.

Defaults

The default of the trigger delay is automatic. 1201 automatically selects

a delay time for you according to the setting of the measurement if you

do not specify a delay. A list of the default for each measurement

function is shown on Table 4-4. The range for the delay is from 0 to

3600 seconds. User's selection of delay time is stored in a volatile

memory and the default will be restored when the meter is power-off.

The internal settings of automatic trigger delay

The delay time for automatic trigger delay is determined by the selection

of measurement function, range, and the setting of integration time and

AC filter speed. The following table shows the default delay time

for each measurement setting.