90

II

II - 2 D

a

ta Int

e

rf

ace

Serial Port

The RS-232-C / V.24 data interface serial port is located on the rear
panel.

The following devices can be connected to this port:

„

Printer with serial data interface.

„

Personal computer with serial data interface.

A printer, or computer may be connected to the serial port. Part
programs, and configuration files may be sent to a printer, or
computer. Remote commands, remote key codes, part programs, and
configuration files may be received from a computer.

For operations that support data transfer, an IMPORT/EXPORT soft
key will be available.

8

The BAUD Field can be set to 300, 600, 1 200, 2 400, 9 600, 19 200
38 400, 57 600, or 115 200 by using the soft keys.

8

The parity can be set to NONE, EVEN, or ODD using the LOWER, or
HIGHER soft key.

8

Data Bits in the Format Field can be set to 7, or 8 using the
soft key provided.

8

The Stop Bits Field can be set to 1, or 2 using soft key.

8

Line Feed Field can be set to YES if the external device needs a line
feed to follow a carriage return.

8

Output tail is the number of carriage returns that will be sent at the
end of the measured value output transmission. The output tail is
initially 0, and can be set to a positive integer value (0 - 9) by using
the numeric hard keys.

The serial port settings will be retained on a power cycle. There is no
parameter to enable, or disable the serial port. Refer to Data Interface
section for cable connection, and pin assignments.