Configuring the compensation table, Automatic non linear error compensation, Ii - 1 installation set u p – ACU-RITE 300S READOUTS User Manual

Page 99

Advertising
background image

300S

85

II - 1 Installation Set

u

p

Configuring the Compensation Table

8

Press the EDIT TABLE soft key to view the table entries.

8

Use the UP, or DOWN arrow keys, or the numeric keys to move the
cursor to the correction point to be added or changed.
Press ENTER.

8

Enter the known error which exists at this point. Press ENTER.

8

When completed, press C key to exit the table and return to the
Error Compensation form.

Reading the Graph

The error compensation table may be viewed in table or graphical
formats. The graph shows a plot of a translation error vs. measured
value. The graph has a fixed scale. As the cursor is moved through the
form, the location of the point on the graph is indicated with a
vertical line.

Viewing the Compesation Table

8

Press the EDIT TABLE soft key.

8

To switch between the table and graph views, press the VIEW
soft key.

8

Press the UP, or DOWN arrow keys or the numeric keys to move
the cursor within the table.

The error compensation table data may be saved to or loaded from a
PC via the Serial port.

Automatic Non Linear error compensation

The correction factor may be calculated automatically using a standard
gauge block. To select automatic calculation, move the cursor to
Spacing, and press ENTER. Press the MANUAL/AUTOMATIC soft
key to select automatic which is now displayed in the Spacing Field.
Move the cursor to any entry within the table (000-199), and
press ENTER.

In the Standard Field, enter the length of the standard plus any tool
offset (tool or edge finder diameter), which is not required when
measuring from the same direction.