Ii - 3 installation and electrical connections, Installation, Electrical requirements – ACU-RITE 300S READOUTS User Manual

Page 105: Environmental, Preventative maintenance

Advertising
background image

300S

91

II - 3 Installation and Electr

ical Connections

II - 3 Installation and Electrical

Connections

Installation

The DRO is mounted to a tilt/swivel feature: See "DRO Dimensions"
on page 104.

Electrical requirements

Voltage

100 - 240 Vac

Power

30 VA max.

Frequency

50/60 Hz (+/- 3Hz)

Fuse 630 mA/250 Vac, 5 mm x 20 mm, Slo-Blo (line and
neutral fused)

Environmental

Operating temperature

0° to 45°C (32° to 113°F)

Storage temperature

-20° to 70°C (-4° to 158°F)

Mechanical weight

2.6 kg (5.2 lb.)

Protective earthing (grounding)

Preventative maintenance

No special preventative maintenance is necessary. For cleaning, wipe
lightly with a dry lint-free cloth.

It is necessary to connect the protective conductor terminal
on the rear panel to the star point of machine ground as
shown in figure to the right.