4.1 Error and special messages

719 S Titrino

103

4 Error messages, Troubleshooting

Data transfer inoperative See measures on page 101.

4.1 Error and special messages

XXX bytes missing

XXX bytes missing. For the storage of a method or a silo line

XXX bytes are missing.

Remedy: <QUIT>. Delete methods no longer needed or use

fewer silo lines.

check electrode

With polarized electrodes. There is a break or short circuit.

Possible causes and rectification of the fault:

- the electrode is not plugged in ⇒ plug it in

- the electrode is not immersed in the solution ⇒ immerse it

- the electrode is defective ⇒ use new electrode.

- the electrode cable is defective ⇒ use new cable.

The electrode test can be switched off under the <PARAM>

key.

Exit: Rectify fault or <STOP>.

check exchange unit

The Exchange Unit is not mounted (properly).

Exit: Mount Exchange Unit (properly) so that the coupling

engages or <STOP>.

check T-sensor

No temperature sensor is attached (with MEAS T or if the

temperature monitoring is active).

Exit: Connect Pt100 or Pt1000 or <STOP>.

division by zero

The result could not be calculated as a divisor in the formula

was equal to zero.

Exit: Enter appropriate value.

manual stop

The determination has been manually stopped.

missing EP

An EP needed for calculation in a formula is missing.

no EP set

In SET, no EP has been set.

Exit: <STOP> and set EP.

no method

The method required by the sample data from the silo memory

or in a TIP sequence is not available in the method memory.

Exit: <CLEAR>.

no new com.var.

The common variable could not be assigned as the result or

the mean value could not be calculated. The old value remains

in force.

no new mean

No new mean value has been calculated as at least one

quantity stipulated for mean value calculations could not be

calculated.

not valid

A value is not available.

overrange

The measuring range of ±2 V has been exceeded. Overrange

replaces the corresponding measured value (pH, U, I or

temperature).

Exit: Rectify error, <STOP> or >MEAS/HOLD>.

same buffer

In the calibration sequence the voltage difference between the

first and second buffer is < 6 mV.

Exit: <QUIT> and change buffer or <STOP> (abort

calibration).