4.2. Diagnosis

719 S Titrino

112

dummy resistor 14.3k ?

• Connect resistor switch-box (or suitable resistor 14.3 kΩ 0.1%) using 3.496.5070

cable to ‘Pol’ socket. Switch-box to 14.3 kΩ.

• Press <Enter>.

Test

sequence:

1. An asterisk flashes during the test.
2. In case of an error an error message appears. (If for example the switch-box is not

connected,

error 100

appears).

3. If no error is found, after about 15 s display shows

polarizer test o.k.

• Press <Clear>.

analog input test 1...5

• Press <Clear>.

diagnose press key 0...9

• Remove cable and resistor switch-box.

8

External inputs and outputs

This test is meaningful only if the 719 S Titrino is used interconnected with other instruments via the 'Remote' connec-

tion. In addition, a 3.496.8510 test plug normally used in the repair service is required for this test. However, this plug

can also be purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the external inputs and outputs

is not required, continue with point 9.

PIN

PIN

PIN

PIN

1  24 5  21
2  12 9  18
3  23 10  17
3  22 11  16

Fig. 3

Connections in the 3.496.8510 test plug

• Press <4>

extern input/output test

• Press <ENTER>.