6.1 Technical specifications

719 S Titrino

127

6 Appendix

6.1 Technical specifications

Modes

SET: Set End point Titration

MEAS: Measurement

CAL: pH calibration

Measuring inputs

2 high-impedance measuring inputs for pH, redox and ISE

electrodes.

1 reference input for a separate reference electrode. May

also be used as a differential amplifier.

1 measuring input for polarized electrodes.

1 measuring input for temperature sensor Pt100 or Pt1000.

Measuring range

pH value (pX)

0...±20.00, resolution 0.01

Voltage

0...±2000 mV, resolution 1 mV, error limit 0.1 % fullscale

Current

0...±200

µA, resolution 1 µA

Temperature

–150.0...+450.0 °C, resolution 1 °C

Polarizer
Ipol

0...±127

µA, resolution 1 µA

Upol

0...±1270 mV, in steps of 10 mV

Measuring amplifier (at 25 °C and Titrino warmed-up)
Input

resistance

>10

13

Offset

current

<3 ∗10

–13

A

Deviation of offset voltage

15 µV/K

Dosification

Volume of buret cylinder 1, (2), 5, 10, 20 or 50 mL

Resolution

10 000 steps per buret cylinder

Titrating burets

1 internal buret

Auxiliary burets

2 additional burets: 776 or 765 Dosimat

Materials

Housing

Polybutyleneterephthalate

(PBTP)

Keypad cover

Polycarbonate (PC)

Display

LCD, 2 lines of 24 characters each

Height of characters 5 mm

LED

back-lit

Internal memory

Method memory for up to 100 methods.

Data bank with 4 Metrohm methods.

Silo memory for sample data and results