7analogue input test – Metrohm 719 S Titrino User Manual

Page 114

Advertising
background image

4.2. Diagnosis

719 S Titrino

110

pot.meter dV/dt  10?

• Turn knob ‘dV/dt’ to the right stop and press <Enter>.

Test

sequence:

a) In a first step, the frequency of the RC oscillator (analogue rate) is tested over a period

of 1 second.

b) In a second step, the frequency of the quartz oscillator (digital rate) is tested over a

period of 1 second.

c) If no error is found, after about 3 s it appears

motor-timer test o.k.

• Press <Clear>.

diagnose press key 0...9

7

Analogue input test

• Press <7>.

analog input test 1...5

7.1

Examination of highly insulated measuring inputs

• Connect ‘Ind I’ measuring input to a voltage calibrator (e.g. Metrohm 773 pH simu-

lator) by means of a highly insulated cable (e.g. 6.2108.060). Set calibrator to 0 V.

• Press <1>.

Input 1

0.0 mV

Tolerance: ± 0.5 mV

• Set the calibrator voltage, on setting 'low ohmic' (with 773 = ~ 0.002 MΩ) to differ-

ent values (e.g. +1500 mV) and compare with the displayed value.

Tolerance (with ± 1500 ÷ 2000 mV) ± 1 mV. Be aware of the calibrator's tolerance.

• Set simulator to high ohmic condition (with 773 = 1000 MΩ).

The displayed reading may vary slightly only (with 1500 mV ≤ 1 mV)

• Press <Clear>.

analog input test 1...5

• Disconnect simulator from ‘Ind I’ input and connect to ‘Ind II’ input.

• Press <2>.

Advertising