2.4

Tutorial

719 S Titrino

13

<ENTER>

<SELECT>

<ENTER>

2 x <QUIT>

Calculate the content of your hydrochloric acid in g/L:

RS1=EP1*C01*C02/C00

End point*conc.(titrant)*molecular weight/sample size

To correct a formula, delete it with <CLEAR>.
• Confirm the formula with <ENTER>.

• You may enter a text for the result output, see page

8.

• Enter the desired number of decimal places for the

result.

• Select the desired unit g/L with <SELECT>, or type

a text as unit, see page 8.

The Titrino prompts for the calculation of the next result.

Quit the formula entry by pressing <QUIT> twice.

In order to be able to calculate the result, enter the calculation constants used in the for-

mula.

Entry of the calculation constants

<C-FMLA>

0.1

<ENTER>

36.47

<ENTER>

• Press <C-FMLA>.

The constants which have been used in the formula

are requested: input with digit keys, confirm with

<ENTER>.

C01: Concentration of your titrant= 0.1 mol/L.

Enter 0.1.

C02: Molar mass of HCl = 36.47 g/mol

Enter 36.47.

Now only the constant C00 is missing to perform the calculation. This constant is re-

served for the sample size. It can be entered directly.

RS1=EP1*C01*C02/C00

>formula

RS1 text

RS1

>formula

RS1 decimal places

2

>formula

RS1 unit

%

RS?

C-fmla

>C01 0.0

C-fmla

>C02 0.0

Advertising