Metrohm 719 S Titrino User Manual

Page 41

Advertising
background image

2.7 Parameters, key <PARAM>

719 S Titrino

37

sample changer cal: OFF







activate pulse: OFF

Calibration with sample changer (ON, OFF)

In calibrations with a sample changer, there are no

hold points in the calibration sequence for inputs.

Calibration temperature and pH values of the buffers

(which are temperature dependent) must therefore be

entered in advance. The inputs in key <PARAM> are

valid.

Pulse output on the line "activate" (L6, pin 1) of the

remote socket (all, first, OFF)

See page 131.

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.