5.1. Setting up and connecting the instruments

719 S Titrino

122

5.1.5 Connection of a recorder

The recorder is connected to the analog output of the Titrino.

The signal at the analogue output can be preselected on the Titrino (key <CONFIG>,

">peripheral units", "curve:"):

Preselection at

Titrino

Meaning Resolution,

Signal at analogue output

U

Voltage

pH = 0.00:

pH = 7.00:

pH = 14.00:

U = + 1 mV:
U = − 1 mV:

I = + 1 µA:
I = − 1 µA:

T = 0 °C:

T = + 1 °C:
T = − 1 °C:

− 700mV

0 mV

+700 mV

+ 1 mV

− 1 mV

+ 10 mV

− 10 mV

0 mV

+ 10 mV

− 10 mV

dU/dt

Measured value drift 1 mV/min:

1 °C/min:

1 µA/min:

1 mV

1 mV

10 mV

V

Volume

1 Zylindervolumen:

2000 mV

dV/dt

Volume drift

100 µL/min:

1000 mV

U(rel) Control

deviation

∆pH = 1:
∆U = 1 mV:
∆I = 1 µA:

100 mV

1 mV

10 mV

T Temperature

∆T = ±1 °C:

T = 0 °C:

10 mV

0 mV

C

B

A

D

719

Recorder