10 common variables – Metrohm 719 S Titrino User Manual

Page 47

Advertising
background image

2.10 Common variables

719 S Titrino

43

2.10 Common variables

Common variables are used for:
• Determination of a titer with a method. This titer is stored permanently as C3X. The

operand C3X can be used in various other methods like any other operand.

• Determination of a blank values with a method . Using this blank value in various other

methods.

• Determination of a result with method. Reconciliation of this result in various other

methods.

You may view the values of the common variables with <CONFIG>.







With <DEF>, results (RSX), endpoints (EPX), variables

(CXX) or mean values (MNX) can be allocated as

common variables.

The entries are specific to the method and are stored in

the method memory.


common variables:
Assigns values as common variables.

The display texts of the Titrino are shown to the left.
The values are the default values.

>common variables

C30=

C31

:

C39=

Allocation for common variables

Common variable C30...C39 (RSX, EPX, CXX, MNX)

Results (RSX), endpoints (EPX), variables (CXX), and

means (MNX) can be assigned.

The values of the common variables remain in force for

all methods until they are overwritten or deleted. They

can be viewed under the <CONFIG> key.

Delete allocation with <CLEAR>+<ENTER>.

DEF

(

2

def

>common variables

Advertising