2.7. Parameters, key <PARAM>

719 S Titrino

28


>stop conditions


stop V: abs.

**titr.






stop V 99.99 ml

**titr.


factor 999999

**titr.



filling rate max. ml/min

Stop conditions for titration

If this is not "normal", i.e. after reaching the EP.

Type of stop volume (abs., rel., OFF)

"abs.":

absolute stop volume in mL.

"rel.":

relative stop volume to sample size.

"OFF":

stop volume switched off. Stop volume

is not monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...±999999)

Calculated as:

Stop V (in mL) = factor ∗sample size

Filling rate after the titration

(0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.