2.8. Result calculations

719 S Titrino

40

RS1 text RS1


RS1 decimal places 2

RS1 unit: %

Text for result output (up to 8 characters)

Text input see page 8.

Number of decimal places for result (0...5)

Selection of result unit (%, ppm, g/L, mg/mL, mol/L,

mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to

6 characters).

Enter next formula, e.g. for RS2.

Meaning of the calculation variables CXX:

C00

Sample size, see page 50.

C01...C19

Method specific operands, see below. They are stored with the method

in the method memory.

C21...C23

Sample specific operands, see page 50ff.

C30...C39 Common

variables.

C40

Initial measured value of the sample, last measured value in MEAS.

C41 End

volume.

C42 Determination

time.

C43

Volume drift for SET with conditioning.

C44 Temperature.

C45

Dispensed start volume.

C46 Asymmetry-pH

(calibration).

C47

Electrode slope (calibration).

Input method specific operands C01...C19, key <C-FMLA>

With <C-FMLA> the operands C01...C19 can be put

in. For the calculation, the operands are used, which

were introduced in the formula.

The inputs method specific and are store in method
memory.

The calculation report can be printed with the key sequence:

<PRINT><SELECT>(press key repeatedly until "calc" appears in the display)<ENTER>